FDA Announces Class I Recall of Tandem Mobi Insulin Pumps: False Alarms Can Stop Insulin Delivery

The U.S. Food and Drug Administration announced on Wednesday, November 25, 2026, that Tandem Diabetes Care is removing certain Tandem Mobi insulin pumps in a Class I recall โ€” the agency's most serious type โ€” because some pumps issue false occlusion alarms and suspend insulin delivery when no blockage exists. The recall covers every Tandem Mobi pump with interoperable technology: pumps on a specific serial-number list must be returned and replaced, while all other users may keep using their pumps under updated instructions. As of October 28, 2026, Tandem had reported 1,248 serious injuries and no deaths associated with the problem, according to the FDA. Here is exactly who is affected and what the FDA says each group should do.

ุงู„ู†ุณุฎุฉ ุงู„ุนุฑุจูŠุฉ

Save

The U.S. Food and Drug Administration (FDA) announced on Wednesday, November 25, 2026, that Tandem Diabetes Care has issued letters telling certain customers to remove Tandem Mobi insulin pumps from where they are used or sold, and that the FDA has identified the recall as its most serious type โ€” a Class I recall, meaning the device may cause serious injury or death if it continues to be used without following the FDA's recommendations. The affected devices are all Tandem Mobi insulin pumps with interoperable technology. One group of pumps, identified by serial number in the FDA's "List of Affected Product Subject to Replacement," must be replaced; a second letter gives updated use instructions to every other Mobi user. Tandem first sent the customer letters on September 23, 2026, and the FDA's recall database entry (Z-0316-2027) covers 20 Mobi models and more than 80,000 pumps, most of them distributed in the United States.

The problem, according to the FDA, is a false alarm. Tandem told the agency that certain Mobi pumps may generate false-positive occlusion alarms when no actual insulin flow obstruction is present, so the pump incorrectly concludes that insulin is blocked and suspends delivery. Delivery stays suspended until the user acknowledges the alarm and resumes therapy. One false alarm is an interruption; repeated false alarms, if not addressed promptly, can interrupt insulin delivery long enough to raise blood glucose levels, and the FDA warns that in severe cases the resulting hyperglycemia may require medical intervention or hospitalization. The Tandem Mobi is a small wearable pump that delivers insulin under the skin for diabetes management and can work with interoperable continuous glucose monitors and automated insulin dosing systems โ€” which is why an unnoticed suspension matters: the system a patient relies on to dose insulin around the clock is the system that has paused.

As of October 28, 2026, Tandem Diabetes had reported 1,248 serious injuries and no deaths associated with this issue, the FDA said. Those figures are the company's reports to the agency, published by the FDA; they are not a finding that every injury was caused by the alarm defect, and the FDA's announcement does not break them down further. What is not in dispute is the mechanism and the remedy, which differ by serial number โ€” so the first step for every Mobi user is to check whether their pump's serial number appears on the replacement list linked from the FDA's announcement.

If your pump's serial number is on the replacement list, the FDA's instructions are direct: confirm the pump in the serial-number look-up list, then immediately submit a request for the return and replacement of your pump, using the link provided in Tandem's customer letter or by calling Tandem Customer Technical Support at 877-801-6901. Until the replacement arrives, always respond to an occlusion alarm and follow the User Guide's troubleshooting steps โ€” acknowledge the alarm, inspect the cartridge, tubing and infusion site for damage or blockage, correct any problem you find, and resume insulin delivery. Make sure the Tandem Mobi mobile app has push notifications enabled so the alarm also reaches your phone, check your blood sugar regularly for unexpectedly high or low readings, and be prepared to use a backup method of insulin delivery as your healthcare provider has directed.

If your pump's serial number is not on the replacement list, the FDA says you may continue to use your pump while following the updated instructions: respond to every occlusion alarm with the same User Guide troubleshooting, keep the app's push notifications enabled, and contact Tandem Customer Support if you begin to experience an unusually high number of occlusion alarms, or alarms that repeatedly disrupt your therapy even after troubleshooting. Keep a backup insulin method ready either way. U.S. customers with adverse reactions, quality problems or questions can reach Tandem at techsupport@tandemdiabetes.com or 877-801-6901, and problems with the device can also be reported to the FDA's MedWatch program. InstaBriefs links the FDA's announcement and the serial-number look-up from this report, and will update it if the FDA changes the scope or the recommendations.

Source attribution

  • FDA โ€” Insulin Pump Recall: Tandem Diabetes Removes Tandem Mobi Insulin Pumps (November 25, 2026)
  • FDA CDRH Recall Database โ€” Class 1 Device Recall, Tandem Mobi, Recall Z-0316-2027 (posted October 30, 2026)
Original source links
  • FDA โ€” Insulin Pump Recall: Tandem Diabetes Removes Tandem Mobi Insulin Pumps (November 25, 2026): Original source
  • FDA CDRH Recall Database โ€” Class 1 Device Recall, Tandem Mobi, Recall Z-0316-2027 (posted October 30, 2026): Original source
About this story

InstaBriefs summary based on the sources listed above โ€” not a copy of the original article.